Major Characteristics for Medical Power Supply

Medical environments are known for its omnipresent risk for both vulnerable and healthy people. To ensure the utmost safety, the design of Medical Power Supplies is monitored by the strictest quality exigences and safety regulations.

As the heart of any electrical medical device, the medical power supplies must comply with the highest standards of performance, safety (IEC 60601) and reliability. Criteria such as MOPP, MOOP, low leakage current, EMC must be taken into consideration during the development stage.

As a result, Medical Power Supplies are equipped with higher levels of insulation and superior EMC performance, and we can distinguish one medically approved Power Supply from other electronics components by its exigent protection level, its high reliability and its extended life cycle of components.

How to choose your Medical Power Supply?

As the keystone of one medical system, it is extremely important to select a safe and reliable Medical Power Supply. To choose the most suitable Medical Power Supply for the Medical system, the R&D engineers are strongly suggested to take, as guideline of selection, the listed criteria below into consideration

 

A. Electrical requirements

 

The input voltage range and required power should be defined for the application, as well as its output voltage and the needed current to power the application:

  • Input voltage range:  the most common input source is an AC source, and the range requirement is related to the destined market of the end devices. In the case of multiple market regions of the Medical system, it is strongly recommended to choose a Medical Power Supply with input voltage range covering at least the range of 100VAC to 240VAC.
  • Power requirement:  the required power depends mainly on the end device. It is suggested to keep in mind the derating caused by the unfavorable working temperature, as well as the potential peak current.

Output voltage and current:  according to the design of the Medical system, some systems demand a constant current output while others need constant voltage. The R&D engineers should choose an adequate Medical Power Supply, with the right level and type of DC output, based on the design of the application.

 

B. Regulations

 

As each market territory and each application field imposes its own safety regulations on the end devices, it is crucial to understand which certifications are mandatory for the application and its target market(s) during the development process.

For instance, home healthcare devices are required to meet the standards of IEC 60601-1-11, while for equipment not directly in contact with the patient, IEC 62368 might still give one the possibility to get a medical certification for medical test and measurement equipment on a system level.

Meanwhile, each country adopts its specific safety standards for the application field (e.g. UL ANSI/AAMI ES60601-1 for the US, EN 60601-1 for EU countries, CAN/CSA-C22 3rd edition for Canada…etc.

 

C. Installation Method

 

The installation of the PSU can be either internally or externally depending on how the Power Supply is designed to be integrated into the system.

External solution:

Internal solution:

  • Medical Enclosed type: PSU with plastic or metal case.
  • Medical PCB type: A PCB populated with components which from the power supply. Normally this solution is without any mechanical protection such as a housing. For this reason, normally installed inside an application and not touchable by the user.
  • Medical On board or Medical PCB mount: PSU designed to fit on a PCB. It can be either “open frame” or “encapsulated”. The second type is potted with compound to provide protection against external elements (e.g. moisture, corrosive elements, shock, vibration…etc.).

 

D. Other requirements

 

Medical systems are regulated with extremely vigorous exigences. On top of the above criteria, it is also important to verify that the other specific requirements are met:

  • Class: Class I or class II
  • Heat dissipation: Forced air cooling (with fan) or Passive cooling (without fan)
  • Level of protection required:  2 x MOOP, 2 x MOPP
  • Leakage current
  • Overload protection: Constant Current Limiting or Hiccup mode
  • Remote on/off function
  • Current sharing function
  • Parallel function

More Information about Medical products?

Specific Requirements for Medical Power Supplies

I. IEC 60601-1:

 

Defined by the world known International Electrotechnical Commission (IEC), IEC 60601-1 is a series of technical standards destined to ensure the safety and performance of electrical equipment in the medical industry. On one hand, some digression from the IEC 60601-1 standards exists. For example, EN 60601-1 for Europe and ES 60601-1 for the US market which are harmonized with the IEC standards. One the other hand, there are also some “collateral” standards to IEC 60601-1; such as IEC 60601-1-2 which is the 4th edition of collateral EMC standards to IEC 60601-1.

 

II. MOOP and MOPP 

 

In the medical environment, risk is constantly present for patient and/or operator of a medical device. Aiming to protect both groups from electrical medical devices, MOOP and MOPP are introduced to monitor the level of insulation.

  • MOOP stands for Means of Operator Protection. This category is related to electronic devices which are handled by trained operators, and do not come into direct contact with the patient. The medical device with 2 x MOOP has an isolation of 3000V AC isolation.
  • MOPP, on the other hand, is the Means of Patient Protection; all electronic devices with direct physical contact with patients must meet stricter standards and require a double isolation between input and output.

Presuming that patients are vulnerable; MOPP standards are specially designed to protect them from any potential electric shock. In the design, MOPP requires the medical devices to have two separate insulation barriers.

Therefore, devices are MOOP and MOPP classified depending on the type of contact with patients and operators.

Classification Required Isolation Required Creepage Requirement Insulation
MOOP 1500V AC 2.5mm
2 x MOOP 3000V AC 5mm Reinforced
MOPP 1500V AC 5mm
2 x MOPP 4000V AC 8mm Reinforced

1 MOOP and MOPP under IEC60601-1 are different in the levels of isolation, creepage, insulation.

III. Low leakage current

 

The essential use of medical equipment is on the human body, therefore the chance and duration of contact with this equipment is higher and longer. Having electric current running throughout the body can be extremely dangerous and might even result in death. For example, a current as low as 40mA could already be fatal to a healthy person, while a weakened person’s tolerance is even lower.

To protect all people in contact with the devices from the electric shock, medical power supplies must meet strict requirements in terms of leakage current. While Applied Parts (AP) indicate the parts of a Medical device or a Medical system which, during normal use, come into direct physical contact with a patient. The standards of IEC60601-1 give a clear definition of the acceptable values of leakage current for each classification of Applied Parts (AP), and the classifications are further divided into 2 categories: “NC” for normal condition and “SFC” for Single fault condition.


Leakage Current
Type B Type BF Type CF
NC SFC NC SFC NC SFC
Earth Leakage Current 5mA 10mA 5mA 10mA 5mA 10mA
Enclosure Leakage Current 100µA 500µA 100µA 500µA 100µA 500µA
Patient Leakage Current 100µA 500µA 100µA 500µA 10µA 50µA

2 Different Leakage Current limits defined by IEC60601-1 according to the medical environment: Type B, Type BF, Type CF

IV. Classification of medical environment

 

Just like medical power supplies, end systems in medical environments are also regulated by strict isolation and leakage current requirements.

Depending on the type of physical contact between patient and medical device, there are three main classification types of Applied Parts (AP) in the medical environment: Type B, Type BF, Type CF.

  • Type B (Body)
    Devices with no direct physical contact with patients. Examples: medical bed, medical laser…etc.
  • Type BF (Body Float)
    Devices with physical contact with patients and might present risk in the case of device failure. Examples: Incubators, diagnostic equipment, etc.
  • Type CF (Cardiac Float)
    Direct contact to the patient’s heart, risk of injury or death in the event of device failure. Examples: Defibrillators, heart-lung machines, etc.

Note: Medical power supplies for type BF & CF devices are designed to meet 2 x MOPP

V. EMC standard and limits

 

Malfunction induced by electromagnetic or interference could be fatal when it comes to life-saving devices. The EMC standards EN 55011 for electromagnetic interference and IEC 60601-1-2 with reference and electromagnetic immunity must be considered.

The 4th edition of the IEC-60601-1 standard on EMC is much more vigorous with electromagnetic immunity than the previous edition. Medical devices must now be immune to HF fields up to 2.7GHz, which represent an increase by 0.2GHz. To prevent damage caused by electrostatic discharge, the limits have also been increased accordingly. For contact discharge, the level has been increased from 6 to 8kV. For air discharge, it has been increased from 8 to 15kV compared to the previous edition.

 

VI. EU Medical Device regulation (MDR)

 

The European Union Medical Device Regulation (MDR) published 2017, will soon replace the current Medical Device Directive (MDD) (93/42/EEC) and the EU’s Directive on active implantable medical devices (90/385/EEC). All development engineers and manufacturers related to Medical Devices within Europe will need to follow the Medical Device Regulation of 2017 published by European Parliament.

 

VII. ISO 13485

 

ISO 13485 is the standard for the quality management system for medical devices. The requirements are destined to organizations involved in the design, production, storage, installation and maintenance of medical devices and other related services.

Got questions?

Look at the section below to find the most frequently asked questions (with answers)
we received in Medical Power Applications.

MEAN WELL’s specification shows the absolute values which were verified during design quality verification tests. Those condition are guaranteed by manufacture from quality and warranty perspectives.

When certifying a power supply according to a certain norm, there is normally a requirement described in this standard a certain tolerance which must be considered. (See also FAQ: Why Is The Input Voltage On The Label Different From The Input Voltage In The Spec? For Example, The Specification Shows Is 88~264 VAC While The Label On The Power Supply Says That It Is 100~240VAC?

The specification shows what is possible, the report and label of the power supply shows what is approved by the certifying body according to the standards.

Besides the difference due to tolerance there might also be another reason why the specification and label/test report show a different temperature. For example, if the power supply needs to be derated at a low voltage input such as 100VAC, the label and test report might show the max temperature at full load based on this low input.  

Different standards might have different tolerance requirements and different ranges this could mean that the most conservative value, or multiple values will show up on the label of the power supply.

In general MEAN WELL’s power supplies with a fan will continue to spin in no load condition. For certain power supplies where the fan will turn off in no load condition this will be marked (1) in the feature section of the specification:

MEAN WELL has incorporated dust proofing and waterproofing into the majority of its LED power supply design. Mainly based on the international standard IEC60529, detailed descriptions can be found in the following table:
(Note: PSUs with IP64 rating or above are suitable for indoor or outdoor applications in sheltered locations)

IP xy protection level

Degree of protection, foreign bodies (x)Degree of protection, moisture(y)
0.    Not protected
1.    Solid foreign object (>50mm)
2.    Solid foreign object (>12mm)
3.    Solid foreign object (>2.5mm)
4.    Solid foreign objects of 1,0mm diameter and greater
5.    Amount of dust that would interfere with normal operation
Dust tight
0.      Not protected
1.      Vertically falling water drop
2.      Vertically falling water drop when enclosure is tilted up to 15 degrees
3.      Water sprayed at an angle up to 60o on either side of the vertical
4.      Water splashed against the component from any direction
5.      Water projected in jets from any direction (12.5 liter/minute)
6.      Water projected in powerfil jets from any direction (100 liter/minute)
7.      Temporary immersion in water ( 1 meeter from the surface of the water for 30 minutes)
8.      Continuous immersion in water, or as specified by the user / manufacture

*IP64-IP66 level products are suitable for damp indoor or sheltered outdoor environment. For actual installation limitations, please refer to the corresponding IP level tests.
*All products cannot be continuously submerged in water.
*The definition of IP68 by MEAN WELL: Immerse a unit under test in 1 meter below the surface of the water, tested with a dynamic condition where 12-hour AC on; 12-hour AC off.
Test duration: 1 month.

Nowadays, customer implement magnetic component in their system to achieve fast installation and maintenance. The magnetic component should keep as far as possible from PSU to avoid interference in the control circuitry of PSU. If limitation of distance is unavoidable, install a well magnetic-conducted metal plate (ex: steel plate, copper plate) between PSU and magnetic component to minimize the interference.

MEAN WELL’s component self-heating can be found on www.meanwell.com

1. Use the search function on the website to find the right product
2. Fill in the series number in the search field (do not include the last extensions such as -12 in XLG-150-12)
3. Click the search button
4. Click on the PDF icon to open the specification

5. Click on report

6. Click on the model and scroll down:

7. The temperature of the critical component can be found in the chapter Reliability Test under item 1 Temperature Rise Test

For more details about the component type for each position, please use the Contact Us function on this website.

MEAN WELL’s DMTBF can be found on www.meanwell.com

Please note that the DMTBF is not for every model available. Mostly this information is available for the LED drivers and our latest product releases.

1. Use the search function on the website to find the right product
2. Fill in the series number in the search field (do not include the last extensions such as -12 in XLG-150-12)
3. Click the search button
4. Click on the PDF icon to open the specification

5. Click on report

6. Click on the model and scroll down:

7. The DMTBF is shown on the last page of the report in the chapter Reliability test

Tag: DMTBF

MEAN WELL’s MTBF can be found on www.meanwell.com

1. Use the search function on the website to find the right product
2. Fill in the series number in the search field (do not include the last 3. extensions such as -12 in XLG-150-12)
3. Click the search button
4. Click on the PDF icon to open the specification

Scroll down in the specification to the bottom of the second page

5. Find the MTBF value in under others:

Tag: MTBF

MEAN WELL’s capacitor lifecycle calculation can be found on www.meanwell.com

1. Use the search function on the website to find the right product
2. Fill in the series number in the search field (do not include the last extensions such as -12 in XLG-150-12)
3. Click the search button 
4. Click on the PDF icon to open the specification

5. Click on report

6. Click on the model and scroll down:

7. The Capacitor life cycle calculation is shown on the last page of the report in the chapter Reliability test

The Capacitor life cycle calculation is considered as the key indicator for the lifetime of the power supply. Please refer to the test report of the power supply on www.meanwell.com  for the capacitor life cycle calculation.

As a rule of thumb, every 10dC increase the lifetime will be cut in half and vice versa for every 10dC decrease in temperature.

In the above example, if the power supply is used at 75dC at Ta 40dC, the estimated lifetime would be 2*104095 Hrs ~200K hrs.

This depends on the condition the power supply is used in. Please contact us for specific situations.
In general, the possibility to use the power supply in a higher temperature than specified is very low and all certifications of the power supply will become invalid.

MEAN WELL offers a selection of Surge Protection Devices to protect the power supply against surges and voltage peaks. 

Go to www.meanwell.com

  1. Select products
  2. Type in “SPD” in the search field
  3. You will find several available SPD’s
Category: Reliability

Most small wattage and fanless power supplies are mainly installed in the horizontal position. If you have to install it vertically because of mechanical limitation, you should consider the output derating due to the heat concern. The temperature derating curve can be found on the spec sheet. Regarding the power supplies with built-in fan or the application has forced cooling system, vertical and horizontal installations have less difference. Ex. In SP-150 derating curve, the ambient temperature difference in application is 5 Celsius from vertical to horizontal. The output wattage in forced cooling can be 20% higher than air cooling convection.

Category: Reliability

Depending on the cooling mechanism of the power supply there are different considerations and restriction towards the mounting orientation of the power supply and distance towards other components and self-heating components.

In general, please check our power supply user manuals on www.meanwell.com (Installation Manual-MEAN WELL Switching Power Supply Manufacturer). The actual distance and orientation depend normally on system design. Customers must review and test the power supplies in the real application and environment.

General guidelines for power supplies with different cooling mechanisms:

  • Power supplies with an internal fan or which are cooled with an external fan have less constraints regarding the orientation in which they are mounted. Nevertheless, the fans and ventilation holes must be kept free from any obstructions. Also, a 10-15 cm clearance must be kept from any adjacent heat source
  • A convection cooled power supply with standard a vertical mounting orientation which will be mounted in another orientation have to be derated by 5dC. See FAQ What should be noticed when installing a power supply in vertical and horizontal directions? Additionally ventilation holes must be kept free from any obstructions. Also, a 10-15 cm clearance must be kept from any adjacent heat source
  • Din rail power supplies can only be mounted in the specified orientation. This is with the ventilation holes at the bottom and on top of the power supply. No other mounting directions are recommended. Allow good ventilation clearances, 5mm left and right, 40mm above and 20mm below, around the unit in use to prevent it from overheating. Also, a 10-15 cm clearance must be kept when the adjacent device is a heat source.
  • For conduction cooling power supplies, the temperature on the Tc point is the key indicator if a mounting position can be allowed. The Tc position (1) and max Tc temperature (2) can be found in the product specification. Conduction cooled power supplies are for example IP65~IP68 LED drivers and base plate cooled power supplies such as UHP)  

In MEAN WELL’s specification you can find 2 derating curves, in below example the 300W open frame power supply: EPP-300

  1. The Derating Curve with the Ambient Temperature vs. Load
  2. The Derating Curve with the Input Voltage vs. Load
  1. In this Curve one can find that the maximum load of this power supply is 300W at 50dC when an external air flow of 20.5CFM is applied. For temperatures exceeding 50dC, 60dC for example additional derating need to be applied, in above case at 60dC the maximum load would be 225W. (50dC 300W, 70dC 150W => from 50dC to 60dC:  75W derating)
  1. In case this power supply would be used at 90VAC input, a derating of 80% must be applied. So, in previous example with the 20.5CFM forced air the max load would be 240W. In case of an ambient temperature of 60dC and a 90VAC input the maximum rated power would be 225W * 0.8 = 180W with 20.5CFM forced air.
  1. If the power supply is used in an application without additional forced Air, the power supply will be derated to 200W till a maximum temperature of 50dC
  2. In case it will be powered by a 90VAC input the power supply has to be additionally derated to 80% of the 200W = 160W max

LED Drivers are recommended operate at full load as long as it observes the working temperature specified in the datasheet, which means Tc measurement results should be equal to or less than the stated Tc in the datasheet. 5 years warranty complies as long as drivers operate within working Temperature and Tc. Limit as well.

MTBF (Mean Time Between Failure) and Life Cycle are both indicators of reliability. MTBF can be calculated by two different methodologies, which are “part count” and “stress analysis”. The regulations, MIL-HDBK-217F Notice 2 and TELCORDIA SR/TR-332(Bellcore) are commonly used to calculate MTBF. MIL-HDBK-217F is a United States military standard, and TELCORDIA SR/TR-332(Bellcore) is a commercial regulation. MEAN WELL utilize MIL-HDBK-217F(Stress Analysis) as the core of MTBF. The exact meaning of MTBF is, after continuously using the power supply for a certain amount of time, the average time that the probability of proper operation is down to 36.8%(e-1=0.368). Currently MEAN WELL is adopting MIL-HDBK-217F, to predict the expected reliability through Stress Analysis (excluding fans); this MTBF means the probability of the product can continue the normal work after working continuously up to the calculated MTBF time is 36.8% (e-1=0.368). If the power supply is continuously used at double the MTBF time, the probability of proper operation becomes 13.5%(e-2=0.135.

DMTBF (Demonstrated Mean Time Between Failures) is a way of evaluating MTBF in a relatively short period of time based on accelerated deterioration (high stress, high temperature) tests under specific parameters and conditions. Please refer to the following equation for MTBF calculation.
This method compares to the previous methods, this calculation uses real test data to demonstrate the reliability of the power supply.

Where

MTBF: Mean Time Between Failure

  • X2:Can be found in chi-square distribution
  • N:Number of sampling
  • AF:Acceleration factor, which can be derived from acceleration factor equation.
  • Ae=0.6
  • K(Boltzmann Constant)=(eV/k)
  • T1:Rated temperature of specification. Note: Kelvin will be the unit use for calculation
  • T2:The temperature that is used in the meaning of acceleration, and the chosen temperature could not result in physical change in materials. 

Note: Kelvin will be the unit use for calculation.  

Life Cycle (Capacitor Life Cycle) is found by using the temperature rise of electrolytic capacitors under maximum operating temperature to estimate the approximate life of the power supply. For example, RSP-750-12 MTBF=109.1K hours(25°C); electrolytic capacitor C110 Life Cycle=213K hours (Ta=50℃).

MEAN WELL considers the capacitor lifecycle calculation as the most important indicator for the estimated lifetime. The (D)MTBF is the main indicator for the reliability of the power supply. For more information please see:  Investigation of The Lifetime & Reliability of Power Supply -MEAN WELL EUROPE Switching Power Supply